LivaNova USA Sterile Disposable Connector,1/2x1/2, 20 units per case

FDA Recall Z-2642-2020 · Posted None

Recalling FirmLivaNova USA Inc.
Recall StatusTerminated
Product CodeDTL
Date Initiated2019-11-18
Date Terminated2021-08-19
Product Quantity140 units

Reason for Recall

Some packages of sterile disposable connectors contained a 3/8x1/4 reducer instead of the specified 1/2x1/2 connector. No significant risk to health is expected as the issue is easily detectable and the different shape would prevent use of the connector.

Firm Action

Affected customers were contacted by letter sent via email or by phone call on November 18th, 2019. Customers were informed that a single lot of LivaNova USA Sterile Disposable Connector item# EC2135S, lot number 1923900091, contained a 3/8 x 1/4 reducer instead of the specified 1/2 x 1/2 connector. Customers were requested to check their inventory, quarantine (and do not use) any affected devices, complete the Customer Response Form and submit the completed form by email to [email protected] to confirm quantity of affected products in the facility. Customers were informed that affected product should be returned to LivaNova, and that their LivaNova reference person would contact customers to organize the replacement of the affected products. Customers were informed that no patient injuries have been reported and that the event is unlikely to cause any harm to the patient or user. Customers were requested to ensure that this Medical Device Correction notice is communicated to all personnel within their organization who need to be aware of it, and, if a customer has transferred any of the affected devices to a third party, to please communicate this information to them and inform the LivaNova Quality Assurance Team at [email protected]. Customers with questions may contact LivaNova at (800) 986-4702 or by e-mail to [email protected]. Customers were informed that this action is being reported to the Food and Drug Administration (FDA). Adverse reactions or quality problems experienced with the use of this product may be reported to LivaNova at [email protected] or the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

Root Cause

Packaging process control

Distribution Pattern

Distributed to medical facilities in 5 states including: Minnesota, Kansas, Missouri, Utah and Washington.