Terminated

Medtronic, Evera Implantable Cardioverter Defibrillators: Product Model EVERA XT DDBB1D1, DDBB1D4, DDBB2D1, DDBB2D4, DVBB2D4, DVBB1D1. EVERA S DDBC3D1, DDBC3D4 Product Usage: The Medtronic Evera DR dual chamber and Evera VR single chamber, implantable cardioverter defibrillator (lCD) are a family of multiprogrammable cardiac devices that monitor and regulate the patients heart rate by providing dual or single chamber rate-responsive bradycardia pacing; ventricular tachyarrhythmia therapies; and atrial tachyarrhythmia therapies (dual-chamber devices only).

FDA Recall Z-2651-2016 · Posted 2016-08-25

Recall StatusTerminated
Product CodeLWS
Date Initiated2016-08-12
Date Terminated2017-01-20
Product Quantity78 (39 US, 39 OUS)

Reason for Recall

78 Viva CRT -Ds and Evera ICDs that were manufactured with a specific subset of circuit components may experience rapid battery depletion due to a low resistance path developing within the circuit component. There have been seven (7) confirmed failures (9%) through July 16, 2016. Reported complications have included shortness of breath, pocket heating, low heart rate, and early device explant.

Firm Action

Consignees were sent or hand delivered on August 12, 2016 a Medtronic "Urgent Medical Device Correction" letter dated August 2016. The letter was addressed to Physician. The letter described the problem and the devices involved in the recall. The letter described a number of "Patient Management Recommendations". For questions contact your local Medtronic Representative or Medtronic Technical Services at 800-723-4636.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution - US Nationwide in the states of US: AZ, CA, FL, GA, IL, KY, MD, MN, NJ, NY, NC, OH, OK, PA, PR, TX. and countries of: Australia, Finland, France, Germany, Iran, Italy, Japan, Korea, Netherlands, Puerto Rico, Saudi Arabia, Spain, and United Kingdom.

Official Source

View FDA Recall →