Terminated

DRX-Revolution Mobile X-Ray System; MODEL DRXR-1; Catalog # 1019397 (all other countries) and Catalog 1060177 (China only) -- COMMON/USUAL NAME: DRX-Revolution --- The DRX-Revolution is a mobile system used to generate and control X-Rays for diagnostic procedures

FDA Recall Z-2651-2017 · Posted 2017-06-19

Recalling FirmCarestream Health Inc
Recall StatusTerminated
Product CodeIZL
Date Initiated2017-05-25
Date Terminated2018-06-13
Product QuantityDomestic: 1449 units; Foreign: 1416 units

Reason for Recall

Carestream found a single unit where four screws in the assembly were not applied in the manufacturing process. This could result in separation of the tube and yoke assembly.

Firm Action

Carestream Health sent the URGENT MEDICAL DEVICE CORRECTION Letters and Response Forms to the US consignees on May 25, 2017, via Certified Mail, Certified Receipt Requested. Letters to foreign accounts were distributed via e-mail or courier about the same time. Customers with questions were instructed to call 1-800-328-2910.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution - US (nationwide) Internationally to Australia, Brazil, Canada, Chile, China, Colombia, Dominican Republic, India, Japan, Israel, Korea, Malaysia, Maldives, Mexico, New Zealand, Peru, Puerto Rico, Singapore, Switzerland, Taipei, Thailand, Turkey, and UAE.