Puritan Bennett Cuff Pressure Manager, REF: 180-03

FDA Recall Z-2651-2023 · Posted 2023-09-29

Recalling FirmCovidien
Recall StatusOpen, Classified
Product CodeBSK
Date Initiated2023-08-22
Product Quantity408

Reason for Recall

Inadequate internal fixation of the power inlet port to the cuff pressure manager device causes the external power inlet to became dislodged within the device housing after several uses, which prevents the device from being charged, and could lead to treatment delay.

Firm Action

On 8/22/23 Medtronic mailed recall notices to risk managers and directors of respiratory care informing them of the following: 1) Discontinue use, quarantine, and return affected devices to the firm. 2) Complete and return the customer confirmation form via email to it to [email protected]. 3) Forward the notice to all those who need to be aware within your organization or to any organization where the potentially affected product with the specified serial numbers has been transferred or distributed. If you have any questions regarding this communication, contact customer service at 800-962-9888, Option 2. Quality problems can be reported to Quality Assurance at 800-255-6774 option 4.

Root Cause

Component design/selection

Distribution Pattern

US: AL, NJ, MA, MI, NY, NC, CO, TX, FL, KY, UT, CA, AZ, KS, IL, TN. OUS: El Salvador, Sweden, Canary Islands