Terminated

Drew Scientific DREW 3-PAC Reagent Kit containing: 5L Diluent, 750 mL Enzymatic Cleaner, and 125 mL Lysing Reagent.

FDA Recall Z-2656-2017 · Posted None

Recalling FirmJAS Diagnostics Inc.
Recall StatusTerminated
Product CodeGKZ
Date Initiated2017-04-21
Date Terminated2020-01-06
Product Quantity14,064

Reason for Recall

May result in elevated platelet backgrounds during routine startup/background checks.

Firm Action

The firm sent notification of recall to consignees on 04/21/2017 by mail in domestic states, and by email or mail to foreign consignees. The notification informed consignees of the product removal. it instructed consignees to discontinue use, destroy product on hand, complete the product destruction form, and contact the firm's customer service department (305-324-2300 or 800-433-0945) to request product replacement.

Root Cause

Material/Component Contamination

Distribution Pattern

AL, AR, CA, FL, GA, IA, IN, LA, MA, MN, MO, NC, NE, NY, SC, TN, TX, and Puerto Rico United Arab Emirates, Bolivia, Republic of C¿te dIvoire, Colombia, Costa Rica, Ecuador, Egypt, Indonesia, Iraq, Italy, Korea, Mexico, Russia, Philippines, Slovenia, Tunisia, Venezuela, and Vietnam

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →