Terminated

ZOLL Pro-padz Sterile Adult Multi-Function Electrodes Part number: 8900-4012 These MFE Electrodes are used in connection with ZOLL Defibrillators for Adult patients. These disposable single use MFE's are processed through Gamma Radiation to a SAL of 1 o.e per International/ National Standards, EN ISO 11137- 2:2012 and ANSI/AAMI/ISO 11137-2:2013, Method 1, for use in a sterile environment by trained medical professionals. The indications for use include defibrillation, cardioversion, external non-invasive pacing and monitoring. These MFE electrodes have a 12 month shelf life from time of manufacture.

FDA Recall Z-2658-2014 · Posted 2014-09-17

Recalling FirmBio-Detek, Inc.
Recall StatusTerminated
Product CodeMKJ
Date Initiated2014-08-26
Date Terminated2016-05-03
Product Quantity198 cases

Reason for Recall

Sterility of device is not assured

Firm Action

BioDetek sent an Urgent Device Correction letter dated August 26, 2014, to all affected customers via UPS overnight delivery. There has also been notifications addressed to the 4 International accounts on August 27, 2014. Customers were requested to return the suspect product(s) in their possession and complete the Device Correction Response Return Form provided to each customer. Customer inventory will be replaced as soon as replacement electrodes are processed. Customers with questions can contact the technical support numbers at 1 (800) 348-9011 or +1 (978) 421-9460 24/7. For questions regarding this recall call 401-729-1400.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution - USA (nationwide) and Internationally to Australia, Canada, and Germany.