Recall StatusTerminated
Date Initiated2016-07-20
Date Terminated2018-07-02
Product Quantity3
Reason for Recall
After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.
Firm Action
Customers were sent a letter dated July 28, 2016 informing them of the problem, under what circumstances it can occur and the actions that the customer can take to minimize the effect of the problem. The letter also includes the actions planned by Philips to correct the problem. Philips plans on releasing a field change order (FCO) to correct the affected systems.
If you need any further information or support concerning this issue, contact your local Philips representative or local Philips Healthcare office. For
North America and Canada, contact the Customer Care Solutions Center
(1 -800-722-9377).
Root Cause
Software Design Change
Distribution Pattern
US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,
MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM,
NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY