Quick Connect 272 KG Scale used with Maxi Sky 2 ceiling lift The Scale is intended to be used with patient lifters to measure the weight of a patient who cannot stand on a scale. It is designed to be used indoors in hospitals, nursing homes, home care or health care facilities.

FDA Recall Z-2660-2018 · Posted None

Recalling FirmArjohuntleigh Magog
Recall StatusTerminated
Product CodeFSA
Date Initiated2018-02-12
Date Terminated2020-08-20
Product Quantity874 Units (both styles)

Reason for Recall

The spreader bar can detach from the scale attachment under certain conditions.

Firm Action

Thefirm, ARJO, sent an "Urgent-A Field Safety Notice" and Customer Response Forms to affected customers by common carrier and signature required. All affected customers need to ensure that all caregivers and users of the affected devices a remade aware of the Field Safety Notice and all listed devices at their facility are available to be corrected during service technician visit. Affected customers need also to review the Field Safety Notice, acknowledge receipt and understanding of the Field Safety Notice and the required actions by completing, signing and returning the Customer Response Form to the local Arjo office. The Arjo Sales and Service office shall arrange Arjo service technician to visit the facilities to carry out the upgrade of the scales attachment on Maxi Sky 2 ceiling lifts. If you have any further questions or require assistance completing the Customer Response Form, please contact Arjo via email at [email protected]

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide distribution: US (nationwide) to states of: CO, FL, IA, IL, KY, MN, NY, PH, and WA; and countries of: Australia, Belgium, Brazil, Canada, China, France, Germany, Hong Kong, Ireland, Italy, Japan, Netherlands, Norway, Singapore, Swiss, Turkey, United Kingdom, and United Arab Emirates.