SafeStep Huber Needle Set with Y Injection Site and Medegen Needleless Injection Cap; Catalog Numbers LH-0029YN (UDI 00801741066177), LH-0030YN (UDI 00801741066191), LH-0031YN (UDI 00801741066214), LH-0032YN (UDI 00801741066238), LH-0033YN (UDI 00801741066252), LH-0034YN (UDI 00801741066276), LH-0035YN (UDI 00801741066290 ), LH-0036YN (UDI 00801741066313), LH-0037YN (UDI 00801741066337), LH-0038YN (UDI 00801741066351), LH-0039YN (UDI 00801741066313), LH-0040YN (UDI 00801741066399)
FDA Recall Z-2669-2020 · Posted 2020-07-20
Advertisement
Recalling FirmBecton Dickinson & Company
Recall StatusTerminated
Product CodePTI
Date Initiated2020-06-19
Date Terminated2022-07-29
Product Quantity5,296,132 total
Reason for Recall
Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
Firm Action
URGENT MEDICAL DEVICE RECALL notification letters dated 6/19/20 were sent to customers.
PLEASE TAKE THE FOLLOWING ACTIONS:
1. Immediately review your inventory for the specific catalog and lot numbers referenced in Attachment A. Discontinue use of those devices and destroy all product subject to this recall following your institutions process for destruction.
MDS-20-1979 Page 2 of 28
BD Medical
1 Becton Drive
Franklin Lakes, NJ 07417
USA
bd.com
2. Share this notification with all users of the product within your facility network to ensure they are also aware of this recall.
3. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your credit accordingly for the discarded material.
a. Customers who purchased product directly from BD will receive credit.
b. Customers who did not purchase product directly from BD will need to contact their distributor to obtain credit.
4. Report any adverse health consequences experienced with the use of this product to BD at [email protected]. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via:
Web: MedWatch website at www.fda.gov/medwatch
Phone: 1800FDA1088 (18003321088)
Mail: MedWatch, HF2, FDA, 5600 Fishers Lane, Rockville, MD 208529787
Customer/Technical Support
Phone: 8002901689
Monday Friday between 8:00 am to 5:00 pm MDT
[email protected]
An Urgent Medical Device Recall notification letter dated 8/25/20 was sent to customers informing them of two additional lots of affected product.
An URGENT MEDICAL DEVICE RECALL UPDATE dated 10/7/20 was sent to customers informing them of two additional lots of affected product.
Root Cause
Process control
Distribution Pattern
Worldwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement