B. Braun Celsite Venous Access Ports , Distributed by Aesculpa Inc. Model ST301, ST305V, ST301G, ST501G and ST315 Reference # Model # 04430425 ST301 04430095 ST305V 04433823 ST301G 04437025 ST501G 04436725 ST315
FDA Recall Z-2671-2011 · Posted 2011-06-27
Recalling FirmB. Braun Interventional Systems
Recall StatusTerminated
Product CodeLJT
Date Initiated2011-04-13
Date Terminated2012-07-06
Product Quantity83
Reason for Recall
The affected products contain the incorrect winged Surecan needle accessory component in certain Celsite Access Port package configurations. The winged Surecan needle is used to deliver medications and fluids into implanted Celsite Access Ports.
Two winged Surecan needle sizes are provided in Celsite U.S. package configurations: 20G (PN 4058631) or 22G (PN 16006593A). A 20G or 22G needle
Firm Action
B.Braun Interventional Systems Inc, (BIS) sent a "Urgent Medical Device Recall" letter on April 16, 2011 to all affected customers.
The letter described the problem and the product involved in the recall. Informed consignees that their local B.Braun Interventional System Inc Product Specialists will contact them with the specifics for return/replacement of the product upon return of qualified removed product. They also requested that a Celsite Access Port Patient Implant Form be filled out for each patient implanted with one of the Celsite Access Ports from the affected lots. This Form to be provided to them by their local BIS Product Specialist.
For additional information contact 800 523-9676 x4459.
Root Cause
Process design
Distribution Pattern
NY, ND, OK ,PA
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K061424CELSITE/ CELSITE CONCEPT, ACCESS PORTSAesculap, Inc. · 2006-07-31K993024CELSITE IMPLANTABLE PORT WITH VALVED CATHETERB. Braun of America, Inc. · 2000-03-13K962230CELSITE PORT WITH PRECONNECTED CATHETERB.Braun Medical, Inc. · 1996-12-16K952548CELSITE VENOUS SYSTEMB.Braun Medical, Inc. · 1995-07-21