ANAX 5.5 Poly-Reduction Screw 6.0X50MM, REF: SFNA6050R, part of ANAX 5.5 SPINAL SYSTEM
FDA Recall Z-2672-2020 · Posted None
Advertisement
Recalling FirmU&I CORP.
Recall StatusTerminated
Product CodeNKB
Date Initiated2020-05-26
Date Terminated2024-04-05
Product Quantity6
Reason for Recall
6.0x50mm poly-reduction screws were laser etched 6.0x45
Firm Action
On 05/26/20, a Request to Return Products Which Have Laser Marking Issue notice was sent to the firm customer. The customer was first asked to distinguish and isolate affected product and then to return affected product. This customer/distributor sent Notification for Sales Suspension and Product Recall notices to their customers. Customers were asked to return affected product and to fill out and return the Notification for Sales Suspension and Product Recall notices to acknowledge receipt.
Root Cause
Process control
Distribution Pattern
U.S.: AZ
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.