Ashvins variant HEYEX 2 / HEYEX PACS. Software Version: Version 2.6.0 (Build 2088) to 2.6.8 (Build 2220). Device Type: Software (Medical Image Management, Picture Archiving and Communication System)

FDA Recall Z-2673-2025 · Posted 2025-09-30

Recall StatusOpen, Classified
Product CodeNFJ
Date Initiated2025-09-09
Product Quantity474 units

Reason for Recall

Potential that the measured value may be smaller than the actual area.

Firm Action

Ashvins notified consignees on about 09/09/2025 via emailed letter. Consignees were instructed to cease using the freeform measurement function in the MMV until the software update has been performed, which is now available. Consignees were instructed to perform the update as soon as possible, complete the digital Field Safety Notice Acknowledgement and Receipt Form, and maintain a copy of the letter and the user manual with the device until the software update is complete.

Root Cause

Software design

Distribution Pattern

US Nationwide distribution.