Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S 4000 TS Tourniquet Systems Single Hose with CPC Connectors Model/Catalog Number: 60-4009-101-00 Software Version: N/A Product Description: Single Sterile Extension Hose with PLC Connector Component: N/A

FDA Recall Z-2674-2026 · Posted 2026-07-02

Recalling FirmZimmer Surgical Inc
Recall StatusOpen, Classified
Product CodeKCY
Date Initiated2026-06-09
Product Quantity14360

Reason for Recall

Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find another readily available device.

Firm Action

On June 8, 2026, the firm began notifying consignees titled "URGENT MEDICAL DEVICE RECALL." Customers were instructed to locate and quarantine affected product in their inventory. Customers may either return the affected product, or a Zimmer Biomet sales representative may remove and return the affected product on the customer's behalf. If the product has been further distributed, the distributor should notify downstream customers and accounts of the recall. All customers should ensure that affected personnel are aware of the recall. This recall is related to prior recall RES 98665, conducted to remove Mixing Bowl and Spatula due to similar issues with seal integrity.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada and EMEA.