Terminated

No packaging. The device is a computed tomography scanner. Intended to provide corss sectional images of the human body and visualization of the internal organs of the body.

FDA Recall Z-2678-2014 · Posted 2014-09-22

Recall StatusTerminated
Product CodeJAX
Date Initiated2007-01-09
Date Terminated2015-11-06
Product Quantity61

Reason for Recall

When the Extended Display Firled ov View reconstruction is used, the CT numbers and anatomy in teh extended region are not accurate.

Firm Action

Philips sent a Product Safety Notification letter dated December 21, 2006 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to distribute the Product Safety Notification to all affected users and keep a copy of the notification with the Instructions for Use until further notice. Customers with questions were instructed to contact Customer Care Center at 1-800-722-9377, option 5 - Diagnostic Imaging, option 1 - CT. For questions regarding this recall call 440-483-7600.

Root Cause

Labeling design

Distribution Pattern

Worldwide Distribution - US including MO, TX, MD, AR, VA, KY, RI, NJ, MA, NY, LA, PA, VT, SC, CA, OH, FL, MN, MI and Internationally to Netherlands, Canada, Finland, Austria, France, Malaysia, Denmark, Norway, UK, Australia, Belgium, New Zealand.