Abiomed REF: 0052-3021, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F, 23F. heart pump and accessories.
FDA Recall Z-2679-2026 · Posted 2026-07-10
Advertisement
Recalling FirmOscor Inc.
Recall StatusOpen, Classified
Product CodeDYB
Date Initiated2026-06-10
Product Quantity1332 units
Reason for Recall
Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region.
Firm Action
On June 10, 2026, Oscor (an Integer subsidiary ) issued a Urgent Medical Device Recall notification to affected consignees. Oscor asked consignees to take the following actions:
1. Product removal is not required, and hospitals may continue to use existing inventory. 2. Physicians should continue to use appropriate clinical awareness, monitoring, and patient support as needed. 3. Consider a device exchange or using the repositioning sheath to minimize blood loss if potential leakage from the sidearm and under the sheath cap and along the hub score lines in the hub region is observed.
4.Review the information within this letter, complete all fields, sign, and return the attached business response form (BRF) (refer to Attachment 1) to [email protected].
5. Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products).
6.If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice.
7. Post a copy of this notice in a visible area for awareness.
Root Cause
Device Design
Distribution Pattern
Distribution in the US state of MA
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source