Terminated

PCR Compano Basic Reader using 100-127V as mains voltage. Model numbers: 732027, 732028, and 732040. Radiological image processing system (Computed Radiography).

FDA Recall Z-2681-2016 · Posted None

Recall StatusTerminated
Product CodeLLZ
Date Initiated2016-08-12
Date Terminated2020-07-15
Product Quantity388

Reason for Recall

A defect component of the Compano reader power supply can cause the power supply to burn and some smoke to leave the housing.

Firm Action

Philips will send a Field Safety Notice to all affected customers on August 12, 2016. The notice informs consignees of the issue and informs them that Philips will replace the power supply unit of the PCR Compano Basic reader system of all affected units free of charge (FCO73200037).

Root Cause

Component design/selection

Distribution Pattern

Worldwide Distribution -- USA, including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OR, PA, TX, VT, WA, WI, WV, and WY; and, the countries of Canada, Brazil, Germany, Iraq, Japan, Korea, Kuwait, Libya, Mexico, Portugal, and Qatar.