Philips DuraDiagnost stationary X-ray system
FDA Recall Z-2682-2016 · Posted 2016-09-01
Advertisement
Recalling FirmPhilips Electronics North America Corporation
Recall StatusTerminated
Product CodeKPR
Date Initiated2015-09-17
Date Terminated2017-08-31
Product Quantity15
Reason for Recall
Buttons or keys on the monitor allow the user to turn the volume setting down to 0", in which case the beep will not be heard at the conclusion of an exposure, as required 21 CFR 1020.31(j), and the Height-Adjustable Tabletop is missing the required labels [21 CFR 1010.2., 21 CFR 1010.3(a), and 21 CFR 1020.30(3)].
Firm Action
Philips plans to bring the products into compliance with the Federal standard at no cost to customers by implementing a field correction.
1. You will contact customers and initiate the actions to correct the failure to comply issues.
2. You will attach the warning label to maintain an audible volume, and issue a supplement to the user manual that instructs the user how to adjust the volume for the local noise environment and include a warning that the user shall maintain an audible volume.
3. You will attach the identification label to the affected Height-Adjustable
Tabletop.
4. The customer notification letter which includes a statement that you will without charge, remedy the defect or bring the product into compliance.
If you need any further information or support concerning this, please contact your local Philips representative or Philips Customer Service: 1-800-722-9377.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
US Distribution to the states of : ME, NE, OR, CA, TX, NC, WA and OH.