Terminated

syngo.via RT Image Suite with software versions syngo.via VB30 or VB40

FDA Recall Z-2688-2020 · Posted None

Recall StatusTerminated
Product CodeMUJ
Date Initiated2020-06-24
Date Terminated2023-04-05
Product Quantity21 units

Reason for Recall

If the user modifies for any reason (e.g. reduction of artifacts) the original image orientation of a standard MR protocol to acquire images in a different orientation for further processing in Synthetic CT , the software does not recognize the adapted acquisition plane. This may result in images with wrong geometry. When this distortion remains unnoticed and the images are subsequently exported to a treatment planning system (TPS), an incorrect calculated radiation treatment plan cannot be excluded. The occurence of this issue is very unlikely and has never been reported so far.

Firm Action

Siemens Healthcare Diagnostics sent a letter notification to customers dated June 23, 2020. The described issue will be solved with a service pack for software versions syngo.via VB30 and syngo.via VB40. Depending on your current software version, the service pack VB30A_HF08 will be installed with update SY039/20/S or the service pack VB40B_HF01 with update SY040/20/S as soon as it is available.

Root Cause

Software design

Distribution Pattern

US Nationwide distribution including in the states of AZ, MA, WI, FL, PA. TX, MD, LA, TX, MN, MI, NY, SC, OH, GA, CA, IL, MO, CA, TX.