Terminated

Software versions Eos software versions: 1.2.4, 1.2.5, 1.2.6,1.2.7,1.2.8, 1.2.9,1.2.10, .2.11,1.2.12, 1.2.13, 1.3.2,1.3.3, 1.3.4, 1.4.0, 1.4.1, 1.4.2. sterEOS software versions: 1.2.3.3788, 1.2.4.4286, 1.3.2.3373, 1.3.3.3671, 1.3.4.3740, 1.3.5.3915, 1.3.6.4272,1.4.1.4813,1.4.2.4850. EOS system is a digital radiography system used in general radiographic examinations. sterEOS workstation is a medical image review station for acceptance, transfer, display, storage, and digital processing of 2D X-ray images.

FDA Recall Z-2692-2011 · Posted 2011-06-28

Recalling FirmEos Imaging
Recall StatusTerminated
Product CodeLLZ
Date Initiated2011-04-13
Date Terminated2012-01-18
Product Quantity5 EOS Systems, 10 sterEOS workstations

Reason for Recall

Print true size, print requests from the EOS acquisition station or the sterEOS image review station may, in certain cases, be affected by a different enlargement factor during printing. T

Firm Action

EOS imaging sent a "FIELD SAFETY NOTICE" letter dated April 13, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to wait for the printing queue to empty before sending a second "print true queue" request with different print settings. Customers should then click the button shown on the instruction sheet to view the print queue. EOS imaging will provide a software update to correct all affected systems at no cost to the customers. An EOS representative will contact customers to schedule a time to upgrade the software. Questions can be answered by calling 33 1 55 25 63 90.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution--USA (nationwide) including the states of CA, IA, MO, VA, and WI, and countries of Canada and Europe.