eye-pak 7407 Tray Support Cover REF 8065740745
FDA Recall Z-2694-2014 · Posted 2014-09-24
Advertisement
Recalling FirmAlcon Research, Ltd.
Recall StatusTerminated
Product CodeHMT
Date Initiated2014-06-03
Date Terminated2015-07-27
Product Quantity2462
Reason for Recall
The peel pouches of the Tray Support Covers may be insufficiently sealed so that sterility cannot be assured.
Firm Action
Alcon Research sent an Urgent Medical Device Recall letter dated June 3, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
Please follow these important steps:
1) Review your inventory of Tray Support Covers to determine if you have any units from the affected Lots in stock.
2) Place the product aside to ensure it is not used.
3) Return the attached Response Form via fax or email to Alcon.
Please return the Response Form even if you do not have any inventory from the affected lots.
4) Alcon will contact you to arrange the return of your inventory.
If you have any further questions, concerns, or need assistance with obtaining replacement product, please contact Alcon Customer Service at 1-800-862-5266.
Root Cause
Packaging process control
Distribution Pattern
Worldwide Distribution - Nationwide Distribution and the countries of Australia, Belgium, and Canada.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.