Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

IntelliVue MX700 patient monitor, model no. 865241

FDA Recall Z-2697-2020 · Posted None

Recall StatusTerminated
Product CodeMHX
Date Initiated2020-06-30
Date Terminated2024-01-25
Product Quantity28

Reason for Recall

Some Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the monitor to stop operating or prevent it from being switched on.

Firm Action

On June 30, 2020, the firm sent Urgent Medical Device Correction letters to affected customers. Customers were instructed to always ensure that the monitor is grounded during operation with mains power. Affected products should be removed from use until the power supply of the monitor has been replaced. Philips Customer Service will contact customers when a replacement power supply is ready for installation. Philips expects all replacements to be complete within six months. If you need any further information or support concerning this issue, please contact your local Philips representative or Philips Customer Support at 1-800- 722-9377.

Root Cause

Process control

Distribution Pattern

Domestic distribution to CA, HI, MA, NY, and UT. International distribution to Costa Rica, Ireland, Lebanon, and Poland.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑