MatrixNEURO Screws - Product Usage: The intended use for this device is in fixation of cranial bones and facial skeleton in procedures such as craniotomies, reconstructive procedures, fracture repair, osteotomies, and selective orthognathic surgery of the maxilla and chin.

FDA Recall Z-2698-2020 · Posted None

Recalling FirmSynthes, Inc.
Recall StatusTerminated
Product CodeJEY
Date Initiated2020-06-18
Date Terminated2021-04-13
Product Quantity104 units

Reason for Recall

One lot of MatrixNEURO 5mm screws was packaged in a body clip etched "4", which indicated a 4mm screw. The outer package labeling correctly identifies a 5mm screw, but the body clop incorrectly identifies a shorter length for the screw.

Firm Action

DePuy Synthes issued urgent field safety notice in June 2020 to notify customers. The affiliates in the impacted countries will notify affected customers.

Root Cause

Under Investigation by firm

Distribution Pattern

International distribution including in the countries of Australia, Belgium, Chile, France, Slovakia, South Korea.