CODMAN ENTERPRISE Vascular Reconstruction Device and Delivery System 4.5mm x 28mm Product code: ENF452812
FDA Recall Z-2701-2011 · Posted 2011-06-29
Advertisement
Recalling FirmCodman & Shurtleff, Inc.
Recall StatusTerminated
Product CodeNJE
Date Initiated2011-03-21
Date Terminated2016-02-17
Product Quantity380 units
Reason for Recall
Mislabeled: One of the stick on stickers on the sterile pouch inside of the product box indicates the stent size to be 22mm in length, when in fact the product is 28mm in length.
Firm Action
Codman & Shurtleff, Inc.sent an "Urgent Medical Device Correction Notice" on March 21, 2011.
The letter recommend the incorrect label be removed and discarded. There are 3 additional peel-off labels, 2 on the outer box and one on the pouch. All 3 of these contain the correct printed size information (28 mm) that can be used for applying to the inventory sheet, patient record or patient card, as needed.
It is NOT required to return any of this product to Codman.
Customer Service at 1-800-225-0460 or contact your local Codman Neurovascular Representative
Root Cause
Labeling design
Distribution Pattern
Worldwide - Nationwide (USA), Canada, Korea.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.