Terminated

Vapotherm Precision Flow,Disposable Patient Circuit PF-NODPC-Low. For use with the Precision Flow system and Ikaria INOmax DSIR System. This DPC allows clinicians to administer nitric oxide non-invasively via high flow nasal cannula. In the Precision Flow High Flow Therapy application, a special disposable patient circuit re-routes the gas to the INOmax injector module then back into the Precision Flow system

FDA Recall Z-2706-2014 · Posted 2014-09-25

Recalling FirmVapotherm, Inc.
Recall StatusTerminated
Product CodeBTT
Date Initiated2014-08-13
Date Terminated2015-10-02
Product Quantity207 units

Reason for Recall

Water may leak into the center gas lumen of the Patient Circuit

Firm Action

Vapotherm issued an Initial Notification on August 13, 2014. The letter was updated on August 20, 2014 to include two additional lots. The recall letter notified users that this defect is easily detectable during normal use and the probability of serious injury or death is remote. The letter was sent to raise awareness of the recent increase in incidence rate of this issue and reminded users of the requirement in the operating manual that requires that the cannula be connected to the patient only after the set temperature is reached. The letter also advised patient care personnel not to use, or to discontinue use of any Disposable Patient Circuit if water is observed leaking into the center lumen or from the cannula. They were instructed to replace the device and report the defect to Vapotherm immediately for further investigation. Should you have any questions please contact our technical support team at Email: [email protected] (US) 855.557.8276 (INTL) 1-603.658.5121

Root Cause

Process control

Distribution Pattern

Worldwide Distribution - USA (nationwide) and Internationally to Canada, France, Swizerland, Korea, Sweden, Germany , Belgium, Spain, Mexico, Scotland, UAE, Australia, Israel, Hong Kong, Ireland, kuwait, and Italy.