Recalling FirmLimacorporate S.p.A
Recall StatusTerminated
Product CodeLXH
Date Initiated2016-07-21
Date Terminated2017-01-13
Product Quantity76
Reason for Recall
Product found to be dimensionally non-conforming (slightly over dimensioned on the diameter of the Zimmer connection) and non-functional for connecting with some T-handles.
Firm Action
The firm received two complaints from the US field about the inta-operative malfunction with the SMR allen wrench 5mm (product code 9013.50.210) with the T-handle. Both parts involved were manufactured with the same lot #14AA311. No adverse effects for patients reported.
Root Cause
Nonconforming Material/Component
Distribution Pattern
US, Italy, Australia
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.