The Schiller PB 1000, also known as Merge Hemo Patient Data Module (PDM), a major component of the Merge Hemo software.
FDA Recall Z-2707-2017 · Posted None
Advertisement
Recalling FirmMerge Healthcare, Inc.
Recall StatusTerminated
Product CodeDQK
Date Initiated2016-03-10
Date Terminated2020-03-06
Product Quantity555 PDM's
Reason for Recall
If the PDM is utilized to output a pressure for computing fractional flow reserve (FFR), the analog output may not reflect the true representation of the selected pressure waveform being provided by the PDM to the Link Assembly which provides the analog output.
Firm Action
The recalling firm issued recall letters dated 3/10/2016, 5/3/2016, and 8/5/2016 via email or via certified mail if the email address is not available.
Root Cause
Software design
Distribution Pattern
Distribution US nationwide.