Terminated

The Schiller PB 1000, also known as Merge Hemo Patient Data Module (PDM), a major component of the Merge Hemo software.

FDA Recall Z-2707-2017 · Posted None

Recall StatusTerminated
Product CodeDQK
Date Initiated2016-03-10
Date Terminated2020-03-06
Product Quantity555 PDM's

Reason for Recall

If the PDM is utilized to output a pressure for computing fractional flow reserve (FFR), the analog output may not reflect the true representation of the selected pressure waveform being provided by the PDM to the Link Assembly which provides the analog output.

Firm Action

The recalling firm issued recall letters dated 3/10/2016, 5/3/2016, and 8/5/2016 via email or via certified mail if the email address is not available.

Root Cause

Software design

Distribution Pattern

Distribution US nationwide.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.