Brilliance CT System 6/10/16/16P/40 configuration. No packaging, sold as single unit. Computed tomography scanner intended to provide cross sectional images of the human body and visualization of the internal organs.

FDA Recall Z-2709-2014 · Posted 2014-09-26

Recall StatusTerminated
Product CodeJAK
Date Initiated2005-12-27
Date Terminated2015-11-06
Product Quantity364

Reason for Recall

The system has a gap that can appear between the upper and lower tilt stand covers. The fingers may be squeezed if a person was to place them in this region while the gantry was tilting from an angle to zero tilt angle.

Firm Action

Customers will be contacted by a field service technician with the field correction order. The firm will install the new labeling to the device.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution -- AZ, CO, GA, IA, IL, IN, KS, KY, MD, MI, MN, MS, NC, NY, OH, OK, TX, VT, and WI; and, countries of Singapore, New Zealand, India, Australia, Korea, Japan, Korea, China, India, Taiwan, Australia, Turkey, France, South Africa, Spain, Italy, Austria, Belgium, Switzerland, Israel, Netherlands, England, and Brazil.