Merge Cardio software. The firm name on the label is Merge Healthcare, Hartland, WI. Image processing system.

FDA Recall Z-2709-2016 · Posted 2016-09-01

Recall StatusTerminated
Product CodeLLZ
Date Initiated2016-02-09
Date Terminated2017-08-24
Product Quantity116 customers potentially have the affected versions

Reason for Recall

A report can be confirmed with the incorrect patient demographics resulting in it being saved under the incorrect patient record in the Electronic Medical Record (EMR).

Firm Action

MERGE sent an Urgent Medical Device Recall letter dated February 9, 2016, to all affected customers via email or via certified mail if they did not have an e-mail address for the customer. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. In addition, the customer is requested to call Customer Service or send them an email to schedule the upgrade. If the customer chooses not to upgrade, they are requested to complete the attached form or go online and complete it. For further questions, please call (800) 724-5970 ext. 8027.

Root Cause

Software design

Distribution Pattern

Nationwide Distribution