Merge Cardio software. The firm name on the label is Merge Healthcare, Hartland, WI. Image processing system.
FDA Recall Z-2709-2016 · Posted 2016-09-01
Advertisement
Recalling FirmMerge Healthcare, Inc.
Recall StatusTerminated
Product CodeLLZ
Date Initiated2016-02-09
Date Terminated2017-08-24
Product Quantity116 customers potentially have the affected versions
Reason for Recall
A report can be confirmed with the incorrect patient demographics resulting in it being saved under the incorrect patient record in the Electronic Medical Record (EMR).
Firm Action
MERGE sent an Urgent Medical Device Recall letter dated February 9, 2016, to all affected customers via email or via certified mail if they did not have an e-mail address for the customer.
The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. In addition, the customer is requested to call Customer Service or send them an email to schedule the upgrade. If the customer chooses not to upgrade, they are requested to complete the attached form or go online and complete it.
For further questions, please call (800) 724-5970 ext. 8027.
Root Cause
Software design
Distribution Pattern
Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.