The 9025TRU Arterial Blood Gas (ABG) Kit is used to obtain arterial blood sampling for pH, Blood Gas, Electrolyte and Metabolite Analysis.
FDA Recall Z-2713-2016 · Posted None
Advertisement
Recalling FirmCardinal Health
Recall StatusTerminated
Product CodeCBT
Date Initiated2016-06-27
Date Terminated2018-03-02
Product QuantityLot #884993 -16,000 units, Lot #903136 -1,900 units and Lot #903137 - 300 units.
Reason for Recall
Arterial Blood Gas (ABG) Kit- Missing Safety Component. The ABG Kit(s) have been identified as missing the locking clasp component. CareFusion/BD has identified a potential customer/user safety risk associated with the missing locking clasp component.
Firm Action
CareFusion/BD sent out a FIELD SAFETY NOTICE of the field correction in writing on June 27, 2016. The letter described the nature of the recall and provided actions to be performed.
Carefusion/BD directed customers to take the following actions:
Step #1: Inspect current inventory on-hand to identify the presence of the affected Arterial Blood Gas Sampling Kit(s).
Step #2: Complete the Customer Response Form by checking all appropriate boxes and providing all accurate information as required.
Step #3: Destroy all affected product(s) in-stock in accordance with their facility's destruction protocol.
Step #4: Return the Customer Response Form to [email protected]
Step #5: If the customers want to obtain credit and/or replacement of destroyed product(s) they are asked to call their CareFusion/BD Customer Support Department at 800.323.9088 (Option #1 ).
For any additional questions and support concerning the voluntary Field Safety Corrective Action, customers are asked to call (847)362.9303.
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide Distribution - US (nationwide) and Canada.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.