Acecide-C High Level Disinfectant & Sterilant 875 mL. Packaging is 6 inner cartons in one master carton;1 inner carton consists of a set of two liquids (1 acid & 1 buffer) banded together.
FDA Recall Z-2714-2016 · Posted 2016-09-02
Recalling FirmBest Sanitizers Inc
Recall StatusTerminated
Product CodeMED
Date Initiated2016-08-02
Date Terminated2017-08-18
Product Quantity3,023 cases USA; 666 cases shipped to Canada (OTC drug there)
Reason for Recall
Certain lots of the acid solution in Acecide-C are degrading faster than normal.
Firm Action
Customer letters dated 8/2/16 will be sent.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Distributed to one (1) account in the state of PA. Distributed to one (1) account in Canada
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.