CelluTome Epidermal Harvesting System Part number CT-H25 or Part Number CT-H50
FDA Recall Z-2717-2014 · Posted 2014-09-29
Advertisement
Recalling FirmKCI USA, Inc.
Recall StatusTerminated
Product CodeGFD
Date Initiated2014-05-19
Date Terminated2015-09-18
Product Quantity1,331 units
Reason for Recall
Blade can shift during shipping which makes the product ineffective.
Firm Action
KCI sent an Urgent - Voluntary Medical Device Field Notification letter dated June 6, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
The firm is sending Field Notifications to affected customers instructing them to complete appropriate actions.
Option I: Contact your local KCI Representative or KCI Customer Technical Support at 1-800-275-4524 and select Option 3 to arrange for return and replacement of the affected harvester(s).
Option 2: If you prefer, you may re-set the blade after removal from the packaging by gently tapping the curved end of the harvester against the palm of your hand, as shown in Figure 3, until the blade is no longer visible in the harvester top plate, as shown in Figure 4.
Root Cause
Device Design
Distribution Pattern
US nationwide Distribution
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.