Cobra, 3-Plate EMG ET Tube, 6.0mm, REF: LTE7003PM, not made with natural rubber latex, Rx Only, STERILEEO
FDA Recall Z-2719-2024 · Posted 2024-08-23
Advertisement
Recalling FirmNeurovision Medical Products Inc
Recall StatusOpen, Classified
Product CodeETN
Date Initiated2024-07-01
Product Quantity5 BOXES (25 single kits)
Reason for Recall
mislabeling; EMG RLN monitoring kit labeling contains the incorrect tube size.
Firm Action
On July 11, 2024, Neurovision Medical Products issued a "Urgent Medical Device Recall Notice" to affected consignees via E-Mail. Neurovision ask consignees to take the following actions:
immediately take the following actions:
1. Review your inventory for the lot mentioned above REF code and lot.
2. Quarantine product subject to this notice.
3. Cease distribution and use of any product from the affected lots.
4. Pass this notice on to all who need to be aware within your organization or to any organization where the potentially affected product with the specified lot number has been transferred or distributed.
5. Complete and return the enclosed response form as soon as possible:
a. Email: customer service @neurovisionmedical.com
b. Fax: (877) 330-1727
6. NMP Customer Service will respond with a return authorization (RMA) number and return shipping label for any remaining stock.
7. If you have distributed this product to another location or entity, notify them of this product recall (include a copy of this notice), and contact NMP Customer Service for return authorization (RMA).
Root Cause
Process control
Distribution Pattern
US Nationwide distribution.