Terminated

Siemens Dimension Vista CHEM 1 CAL, lot 3GM081. The CHEM 1 CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyroxine Uptake (TU), Blood Urea Nitrogen (BUN), and Uric Acid (UA) methods on the Dimension Vista System.

FDA Recall Z-2720-2014 · Posted 2014-09-30

Recall StatusTerminated
Product CodeJIX
Date Initiated2014-08-18
Date Terminated2015-02-02
Product Quantity9769 Total (6000 US)

Reason for Recall

Siemens Healthcare Diagnostics has received complaints for low recovery of Magnesium (MG) Quality Control (QC) and patient samples following calibration of MG with the Dimension Vista® CHEM 1 CAL, lot 3GM081. Investigation by Siemens confirms that QC and patient samples under-recovers MG by 0.25 mg/dL [0.10 mmol/L] across the assay range following calibration with 3GM081.

Firm Action

Siemens sent an Urgent Medical Device Dorrection letter, dated August 2014, to all affected consignees. The letter identified the product, the problem, and the actions to be taken. Consignees were instructed if they still have an active calibration of MG with this lot of CHEM 1 CAL, they should immediately calibrate MG with an alternate lot of CHEM 1 CAL. Consigness were instructed to retain the letter with their laboratory records, and to forward the letter to those who may have received the affected product. Customers with questions were instructed to contact their Siemens Customer Care Center - Techhnical Solutions or their local Siemens technical support representative.

Root Cause

Process design