ClearPro Trach T-Piece Closed Suction Catheter, Model DYNCPTP14T, 14 Fr 12.5 IN FOR TRACHEOSTOMY TUBE - Product Usage: Catheters are intended for endotracheal suction of adult patients with artificial airways. Single use only.

FDA Recall Z-2725-2020 · Posted None

Recall StatusOpen, Classified
Product CodeBSY
Date Initiated2020-06-30
Product Quantity680 units

Reason for Recall

There are reports of the suction catheter coming apart from the device assembly during use.

Firm Action

Customer letter dated 06/29/2020 was issued to consignees. Consignees are asked to check their inventory for impacted product and destroy affected product. Consignees are asked to complete and return a destruction form listing quantity of affected product destroyed or indicate no product was on hand. Additionally, if product has been further distributed, consignees are asked to notify those customers who may be impacted.

Root Cause

Process control

Distribution Pattern

US Nationwide distribution including in the states of AR, AZ, CA, FL, GA, IL, MS, NY, NV, OH, PA, TX.