Terminated

Access Thyroglobulin Assay; 100 determinations, 2 packs, 50 tests/pack. For the quantitative determination of thyroglobulin levels in human serum and plasma using the Access Immunoassay Systems.

FDA Recall Z-2733-2015 · Posted 2015-09-03

Recalling FirmBeckman Coulter Inc.
Recall StatusTerminated
Product CodeMSW
Date Initiated2015-07-28
Date Terminated2016-04-13
Product Quantity3,035 (U.S.) and 7,123 (International)

Reason for Recall

Beckman Coulter, Inc. is recalling Access Thyroglobulin reagent packs because some may exhibit greater than 10% imprecision at concentrations <10 ng/mL.

Firm Action

On 07/28/2015 the firm sent out customer notification letters via first class mail and e-mail. The firm's letter contained the following instructions: 1. Discard all Access Thyroglobulin reagent pack lots listed in the letter. 2. Evaluate Access Thyroglobulin results <10 ng/mL when using the reagent pack lot numbers listed to determine if controls and/or patients were higher than expected or did not match the clinical status of the patient. 3. At the discretion of the Laboratory Director, notify clinicians that it is possible the Access Thyroglobulin sample results reported by the laboratory were affected by this issue. The firm's notification letter requests that the information be shared with laboratory staff. If any affected products have been forwarded provide a copy of the letter to them. The firm requests that customers complete and return the Response Form within 10 days. Any questions contact the Customer Care Center: 1. From website at http://www.beckmancoulter.com 2. By phone: call 1-800-854-3633 in the United States and Canada. 3. Outside the United States and Canada, contact local Beckman Coulter representative

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Algeria, Australia, Austria, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Korea, Republic of, Lebanon, Lithuania, Mexico, Morocco, New Zealand, Oman, Philippines, Portugal, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, Vietnam and Yemen.