Integra Cranial Access Kits, Catalogue Nos. INSHITH; INSHITHRZN; INSHITHND
FDA Recall Z-2733-2016 · Posted 2016-09-02
Advertisement
Recall StatusTerminated
Product CodeHBG
Date Initiated2016-08-24
Date Terminated2017-03-30
Product Quantity710
Reason for Recall
Integra LifeSciences has been notified of a medical device recall by Adaptive Surgical, LLC involving their Riverlon" Nylon 3-0 Sutures that are packaged in Integras Cranial Access Kits. It was found that product seals may become open during extreme distribution conditions and render the device unsterile.
Firm Action
Consignees notified on August 12, 2016 via letter to identify and return product.
Root Cause
Packaging
Distribution Pattern
Nationwide and VA/govt/military. No foreign consignees.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.