2 over 2 Elevating Monitor Suspension - Omega Medical Imaging, LLC Elevating Monitor Suspension Systems.

FDA Recall Z-2734-2016 · Posted None

Recall StatusTerminated
Product CodeN/A
Date Initiated2016-06-21
Date Terminated2019-10-04
Product Quantity26 units

Reason for Recall

Elevating Monitor Suspension separated from the lifting column bracket

Firm Action

Consignees were notified by a Field Safety Advisory Notice - FSAN 16-0001 of this immediate Recall on 6/21/2016.

Root Cause

Device Design

Distribution Pattern

CO, MN, GA, NC, IN, OH, KY, IL, NY, AL, MO, LA, PA, and SD Denmark