Open, Classified

Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation.

FDA Recall Z-2735-2026 · Posted 2026-07-28

Recalling FirmResMed Ltd.
Recall StatusOpen, Classified
Product CodeCBK
Date Initiated2026-06-25
Product Quantity168,069 devices globally

Reason for Recall

Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.

Firm Action

ResMed Urgent Medical Device Correction Astral 100/150 Ventilators Issue Identified: On June 25, 2026, ResMed notified customers via Salesforce of an urgent medical device correction involving a subset of Astral 100 and Astral 150 ventilators. An internal supercapacitor may leak electrolyte over time, potentially damaging the printed circuit board assembly (PCBA) and causing the ventilator to enter a fail-safe state. Interim Guidance: Due to limited PCBA replacement availability, immediate correction of all affected units is not feasible. Physicians and patients are advised to follow Astral 100/150 User and Clinical Guides, including recommendations on monitoring, emergency preparedness, caregiver training, and backup ventilation. Required Actions for Healthcare Providers and Distributors: Acknowledge: Complete and return the acknowledgment form by July 31, 2026. Notify: Distribute the correction notice and accompanying letters to all relevant healthcare professionals, patients, and caregivers immediately. Reinforce Safety Measures: Ensure ventilator-dependent patients are monitored, caregivers are trained in alarm response and emergency procedures, and backup ventilation equipment is functional and readily available. Device Management: Do not remove an affected ventilator from use unless an alternative is immediately available. Replace inoperable units promptly; patients should continue therapy unless directed otherwise by their clinician. Risk Assessment: Evaluate affected patients by clinical risk category (Tier 1, 2, or 3) per Appendix A. Arrange Inspection: Schedule ventilator inspections per the Phase 1 Inspection and Correction Strategy in Appendix A, returning units to an authorized service center as needed. Contact: Reach your local ResMed representative or visit www.resmed.com/contact for questions.

Root Cause

Component design/selection

Distribution Pattern

Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Great Britain, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Islamic Republic of Iran, Israel, Italy, Japan, Jordan, Kenya, Republic of Korea, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Saint Martin (French part), Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Uruguay, and Viet Nam.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →