Terminated

VitalGo Total Lift Bed VG-TLB425T-TQR, Version 71064V0-27003

FDA Recall Z-2739-2018 · Posted None

Recalling FirmVitalgo Inc.
Recall StatusTerminated
Product CodeFNL
Date Initiated2018-04-12
Date Terminated2019-03-12
Product Quantity85 units

Reason for Recall

Potential for unintentional movement of the support surface of the bed from tilted position to horizontal position or any position in between.

Firm Action

The customer was notified of the recall on about 04/12/2018 via letter. The letter was follow up to previous communications on 02/19/2018, 03/06/2018, 03/07/2018, 03/26/2018, and 04/09/2018 and reminded the customer that all affected beds at each service center are to be placed on Quality Hold. Further directions will be provided at a later time to conduct a correction on the beds.

Root Cause

Component change control

Distribution Pattern

US distribution to Texas.

Official Source

View FDA Recall →