Slit Knife 2.6mm 45degree Bevel Up, Model Number 373726. for use in ophthalmic surgery.
FDA Recall Z-2740-2018 · Posted None
Advertisement
Recalling FirmBeaver Visitec
Recall StatusTerminated
Product CodeHNN
Date Initiated2018-06-18
Date Terminated2019-10-30
Product Quantity440
Reason for Recall
Mislabel of product 373726 Lot 6001700 with 374891 Lot 6001701 on individual packages within the correctly labeled shelf carton.
Firm Action
BVI issued PRODUCT RECALL NOTICES to customers dated June 11, 2018. Customers were instructed to inspect stock, quarantine and destroy product per internal procedures. BVI is kindly asking the customers to provide objective evidence of product destruction so that a credit can be applied to accounts. Customers are also encouraged to complete and return the response form. If you have questions regarding this recall please call Customer Service at 866-906-8080.
Root Cause
Packaging process control
Distribution Pattern
Distribution to US states of CA, IL, MD, MN, NJ, OH & PA; and Israel
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.