CPAP Mask Cushion, Model Nos. PB781S, PB781M, PB781L, 781S, 781M and 781L Product Usage: A CPAP mask is attached to an end user s face and a CPAP machine, to treat end users with sleep apnea. The replacement cushion is a non-durable component that requires replacement every 3 - 6 months. The cushion is used for both comfort and to provide a seal between the face and the mask.
FDA Recall Z-2741-2018 · Posted 2018-08-20
Advertisement
Recalling FirmCompass Health Brands (Corporate Office)
Recall StatusTerminated
Product CodeBZD
Date Initiated2018-06-18
Date Terminated2020-05-05
Product Quantity742
Reason for Recall
A May 2015 design change in the cushion seal replacement part and accompanying elbow replacement part resulted in incompatibility of the new cushion seal with the previous design of the elbow. Use of the new cushion seal with the previous design elbow could lead to unacceptable CO2 porting. As there was no model number change for this design change, customers may not be aware of the incompatibility.
Firm Action
On June 19. 2018, the firm sent Urgent Recall Field Correction notices via certified mail. The notice alerted customers to the following product issue:
Compass Health Brands Corp. has recently identified the need to clarify
compatibility information for the replacement cushion seals for the Model# PB7801 Zzz-Mask SG Series Full Face CPAP Mask.
>- A design change made to the cushion seal replacement part (and accompanying elbow
replacement part) in May 2015 (18 months prior to Compass Health acquisition of the Pro basics
product line) resulted in incompatibility of the new cushion seal with the previous design of the
elbow.
>- While no complaints or injuries have been reported to-date, use of the new cushion seal with
the previous design of elbow could result in unacceptable CO2 porting.
>- Since a model number change did not accompany this design change, customers and their
consumers would likely not be aware of this incompatibility.
Customers were asked to do the following:
Ensure all affected personnel are fully informed of this notice.
" Forward this notice to your Regulatory Compliance Manager, Purchasing Manager, Customer
Service Manager and Field Technicians.
" Advise your customers regarding the proper combination of cushion and elbow. The attached
consumer communication letter can be used to communicate this information. As noted on the
attached consumer communication letter, any consumers who are currently using an
incompatible cushion/elbow combination should be instructed to discontinue use of the
cushion and elbow and should be instructed to contact their dealer to obtain replacement kits.
" Dispose of any current stock you may have of the affected cushion model numbers (PB781S,
PB781M, PB781L, 781S, 781M and 781L manufactured after May 2015) and contact Compass
Health for replacement kits (NOTE: Full mask model numbers are NOT affected by this recall
field correction).
" Please contact Compass Health Brands Corp. Customer S
Root Cause
Component change control
Distribution Pattern
US nationwide distribution to AL, CT, FL, GA, KY, MS, NC, NY, TN, and TX.