The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II device that was cleared for marketing by FDA under 510(k) number K101274.
FDA Recall Z-2741-2026 · Posted 2026-07-15
Advertisement
Recalling FirmMaquet Cardiovascular, LLC
Recall StatusOpen, Classified
Product CodeGEI
Date Initiated2026-06-02
Product Quantity205 units
Reason for Recall
Getinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma sterilization run (Run ID 1183-28782A) performed by STERIS at its Whippany, NJ facility between March 31 and April 1 of 2026.
Firm Action
Getinge issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 6/18/2026 via FedEx. The notice explained the issue, risk to health, and requested the following actions be taken:
" Immediately identify and quarantine all affected Vasoview Hemopro 2 devices from lot 3000547797.
" Do not use affected devices.
" If an affected device was used, no further intervention is required. Standard post-operative wound care per facility protocol should be followed.
" If any affected devices were transferred within your organization or to another organization, ensure this notice is shared with the appropriate individuals.
" Contact Getinge Customer Service 1-888-880-2874 or your Getinge representative to coordinate return of the affected units and replacement product.
" Complete and return the enclosed response form to Getinge by e-mailing a copy to [email protected].
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide distribution in the states of NY, CA, SC, FL, TX, GA, OR.