Cambridge Endoscopic Autonomy Laparo-Angle 5mm Metzenbaum Scissors with Monopolar Cautery, 45 cm Catalog Number: LA1101-12 Intended for use in a variety of minimally invasive procedures to facilitate grasping, mobilization, dissection, and transection of tissue.

FDA Recall Z-2744-2011 · Posted 2011-07-06

Recall StatusTerminated
Product CodeGEI
Date Initiated2011-05-20
Date Terminated2012-07-24
Product Quantity255 units

Reason for Recall

PET shrink sleeve insulation material dislodging from the distal tips of Autonomy Laparo-Angle (ALA) Metzenbaum Scissors

Firm Action

Cambridge Endoscopic Devices, Inc. sent an "IMPORTANT MEDICAL DEVICE RECALL" letter dated May 20, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. Customers are instructed to return unused inventory using RGA Number 301. Returned devices will be replaced. Questions regarding this recall, call 508-405-0790 for assistance.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution--USA (nationwide) and the countries of Armenia , Australia, Austria, Chile, China, Denmark, Hong Kong, India, Ireland , Italy, Japan, Saudi Arabia, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.