BD 3ml Syringe Luer-Lok" Tip with BD PrecisionGlide" Needle 23G x 1 (0.6mm x 25mm) Catalog #309571 The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection

FDA Recall Z-2749-2017 · Posted None

Recall StatusTerminated
Product CodeFMF
Date Initiated2016-09-16
Date Terminated2018-03-20
Product Quantity57600 units

Reason for Recall

BD identified that several lots did not receive the specified dose of irradiation necessary to meet the Sterility Assurance Level (SAL) required.

Firm Action

BD sent recall letter/return response form on 9/16/2016.

Root Cause

Process control

Distribution Pattern

Nationwide Distribution