BD 3ml Syringe Luer-Lok" Tip with BD PrecisionGlide" Needle 18G x 1 1/2 (1.2mm x 40mm) Catalog #309580 The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection
FDA Recall Z-2750-2017 · Posted None
Advertisement
Recalling FirmBecton Dickinson & Company
Recall StatusTerminated
Product CodeFMF
Date Initiated2016-09-16
Date Terminated2018-03-20
Product Quantity340,000 units
Reason for Recall
BD identified that several lots did not receive the specified dose of irradiation necessary to meet the Sterility Assurance Level (SAL) required.
Firm Action
BD sent recall letter/return response form on 9/16/2016.
Root Cause
Process control
Distribution Pattern
Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.