Terminated

Monitor with one Blood Parameter Module and one HSat Probe; 500AHCT Monitor with two Blood Parameter modules and one HSat Probe; 500AVHCT Product Usage: The CDI System 500 provides continuous, on-line monitoring of the extracorporeal partial pressure of oxygen and carbon dioxide, pH, potassium, oxygen saturation, hematocrit, hemoglobin and temperature. In addition, calculated values of base excess, bicarbonate, oxygen saturation, and oxygen consumption may also be provided. These parameters are displayed at either actual temperature or adjusted to 37C. For documentation purposes, the system 500s integral printer provides a hard copy of displayed parameters.

FDA Recall Z-2751-2018 · Posted None

Recall StatusTerminated
Product CodeDRY
Date Initiated2018-06-26
Date Terminated2020-04-17
Product Quantity128 in total

Reason for Recall

H/Sat Hybrid Intensity Shift has been seen in a subset of the H/Sat population, which has resulted in an increase of Hybrids failing the H/Sat Color Chip Test which is an automated self-test when the device is powered on. This will not result in device inaccuracy, only Color Chip Test Failure which disables the H/Sat module.

Firm Action

On June 26, 2018, the firm began notifying customers of the market withdrawal via telephone calls. The call alerted customers of the issue and scheduled an expedited return of affected CDI System 500 units to the firm's service depot so that Service Technicians may replace the affected H/SAT with a replacement H/SAT device. Customers then received a copy of the completed phone script by email along with a customer response form. Customers were asked to review the phone script, and complete and return the response form. Customers were also asked to assure that all users were notified of the issue, and to prominently display the notice where all users may access it. To contact Terumo CVS with any questions or concerns: Terumo CVS Customer Service: 1.800.521.2818 (Monday Friday, 8 a.m. 6 p.m. ET) Recall Fax: 1.734.741.6149

Root Cause

Vendor change control

Distribution Pattern

Worldwide Distribution - US Nationwide in the states of AL, AZ, CA, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, MS, NC, NM, NY, OH, SC, TN, TX, WA, and WV. Foreign distribution to Middle East, Latin America, Europe, Japan, Canada, and Mexico.

Official Source

View FDA Recall →