FARASTAR Pulsed Field Ablation Generator, Product ID M004PFCE61M401; Percutaneous Cardiac Ablation Catheter
FDA Recall Z-2751-2026 · Posted 2026-07-17
Advertisement
Recalling FirmBoston Scientific Corporation
Recall StatusOpen, Classified
Product CodeQZI
Date Initiated2026-06-11
Product Quantity1 units
Reason for Recall
Labeling discrepancy: Devices distributed with Instructions For Use (IFU) did not include
FARAPOINT PFA catheter information.
Firm Action
Boston Scientific issued an Urgent Field Safety Notice - Urgent Medical Device Correction notice to its consignees out side the US on 6/11/2026 via hand delivered letter. The notice explained the issue and requested the following actions be taken:
Instructions:
1- Post a copy of the appropriate IFU provided near each FARASTAR PFA generator, replacing any previous versions currently in use.
2- Review the IFU and ensure all relevant personnel are aware of its contents.
3- Please complete the attached Acknowledgement Form even if you do not have any affected product and return it to your Boston Scientific office.
4- Share this communication with any healthcare professionals in your facility that use the product and with any other organization to which this product may have been transferred.
The national Competent Authority was informed of this communication. Any adverse events or quality concerns associated with use of these devices should be reported to Boston Scientific and Competent Authorities if appropriate.
Root Cause
Under Investigation by firm
Distribution Pattern
International distribution in the states of Austria, Spain, France, Germany, Denmark, Italy, Portugal, Great Britain, Ireland.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.