Open, Classified

FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21M402, IFU Reference Number LBL1062

FDA Recall Z-2755-2026 · Posted 2026-07-20

Recall StatusOpen, Classified
Product CodeDRA
Date Initiated2026-06-11
Product Quantity74,760 units

Reason for Recall

Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.

Firm Action

An Important Medical Device Product Advisory notification letter dated 6/11/26 was sent to customers. Actions " Follow the revised IFU provided in Appendix 1. " Forward this notice to relevant personnel and any facilities to which the product was transferred. " Complete and return the enclosed Acknowledgement Form. " Report any adverse events or quality concerns associated with use of these devices to Boston Scientific via email at [email protected] or FDA s MedWatch Adverse Event Reporting program [www.fda.gov/MedWatch/report.htm or 1.800.FDA.1088 (332.1088)]. If you require additional assistance or more information regarding this communication, please contact your local Boston Scientific representative.

Root Cause

Use error

Distribution Pattern

Global Distribution. US Nationwide.