Terminated

UCLA Abutment Hexed Castable Cylinder 3.4mm Are accessories to endosseous dental implant to support a prosthetic device in a partially or completely edentulous patient.

FDA Recall Z-2760-2015 · Posted 2015-09-11

Recalling FirmBiomet 3i, LLC
Recall StatusTerminated
Product CodeNHA
Date Initiated2015-07-29
Date Terminated2016-05-16
Product Quantity61 pieces

Reason for Recall

MUCC2, non-hexed, castable abutment cylinders, has been labeled and distributed as a MUCC1C, hexed, castable abutment cylinders for this one lot only.

Firm Action

Biomet sent a Medical Device Recall Notice dated July 27, 2015, to all their clinicians informing them about the recall and to complete the attached recall return response form and any unused product to be replaced at no charge. Contact information 1-800-342-5454

Root Cause

Process control

Distribution Pattern

Worldwide Distribution including Italy and Uruguay.