Cascade Abrazo aPTT Test Card, Model 5722, For Export Only. The firm name on the label is Helena Laboratories, Beaumont, TX.
FDA Recall Z-2761-2019 · Posted 2019-09-27
Advertisement
Recalling FirmHelena Laboratories, Corp.
Recall StatusTerminated
Product CodeGFO
Date Initiated2017-12-07
Date Terminated2020-06-09
Product Quantity21 kits
Reason for Recall
Some of the cards show occasional shorter clot times than the expected variation resulting in a potential false negative result on the screening assay.
Firm Action
The recalling firm issued letters dated 11/12/2017 via email on 12/7/2017 explaining the issue and that they were doing a withdrawal of the lot.
Root Cause
Unknown/Undetermined by firm
Distribution Pattern
Distribution was made to Germany and Australia. There was no U.S. distribution.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.